With ICP MS, whether all elements present in the Periodic table can be analysed or not, then what are all the challenging elements by ICPMS?

user-image
Question added by Arivumani Durairaj Servai , Sr. Engineer/ Asst. Manager , Emphor, Centena group, Life & Analytical Science Instruments Division
Date Posted: 2014/07/17
Elamin Osman Elamin Mohammed
by Elamin Osman Elamin Mohammed , Technical Manager , SAMA OIL INDUSTRIAL CO,

ICP MS is able to detect all metals and several of non-metal with very low concentration ,because the ions formed by the ICP discharge are typically positive ions, M+ or M+², therefore, elements that prefer to form negative ions, such as Cl, I, F, etc., are very difficult to determine via ICP-MS.

Mahesh Selvarajan
by Mahesh Selvarajan , Quality Manager , Kuwait

Most of the elements in the periodic table can be analysed using ICP-OES or ICP MS. But some elements especially gases cannot be analysed. 

Follow this link to get an idea of detection limit.

http://www.eag.com/documents/icp-oes-ms-detection-limit-guidance-BR023.pdf

Arivumani Durairaj Servai
by Arivumani Durairaj Servai , Sr. Engineer/ Asst. Manager , Emphor, Centena group, Life & Analytical Science Instruments Division

Most of the elements can be done except few elements and gases. By ICPMS - Halogens are difficult to analyze.